Creating a Processing Sequence
Editing a Processing Sequence
Resolving Discrepancies in Level Names
When you add samples to a sequence, if the LCquan application detects a discrepancy in the
level names, one of these dialog boxes opens:
• The Standard or QC Level Names Association dialog box opens when the samples are
associated with calibration or QC level names that do not match the level names of the
workbook sequence. For more information, see “Resolving Mismatched Level Names” on
page 319.
• Workbook Does Not Contain Levels dialog box opens when the samples are associated
with level names, but the workbook processing method has no calibration or QC levels
defined. For more information, see “Resolving Missing Calibration or QC Levels” on
page 321.
Resolving Mismatched Level Names
The Standard (or QC) Level Names Association dialog box warns you of a discrepancy in the
calibration level names or QC level names under the following conditions:
• Importing an acquisition sequence—The level names in the acquisition sequence do not
match the level names in the workbook.
• Importing a processing sequence—The level names in the imported processing sequence
do not match the level names in the workbook processing method.
• Adding sample files to a processing sequence—The level names associated with the added
samples do not match the level names in the workbook.
Y To resolve the level names for an imported sequence
1. Verify that you imported the correct sequence.
2. To associate the level names, do one of the following (Figure 186):
• Select the level name in the Sequence Level Names pane and drag it to the
appropriate level name in the Method Level Names pane (processing method) or the
Workbook Level Names pane.
–or–
• Select the two level names and click Add.