Type 8045
Operating and functions
→ Charge the pipe.
→ Stop the fluid circulation.
→ Make sure the fluid is still.
CALIB. 0
CALIB N
0......9
CALIB Y
→ Choose "CALIB Y".
MEASURE /
The device calibrates the flow zero point.
FLOW
The calibration has
succeeded.
CAL. FAIL
ENTER
The device has detected a non nil
flow rate in the pipe.
The device status LED becomes
orange.
Fig. 58:
Flow zero point calibration using the "CALIB 0" parameter of the Test menu
→ If you do not want to adjust another parameter, go to the "END" parameter of the Test menu and press
ENTER
to save the settings or not and go back to the Process level.
8.7.3.
Checking the outputs behaviour
This parameter makes it possible to check that the outputs are behaving as expected by the configuration made.
• The totalizers are incremented depending on the measured value of the flow rate and not the simulated
value.
• The device status LED flashes during the running check of the output behaviour.