Calibration of Oxygen-Sensitive Microsensors 68 5. Enter the atmospheric pressure at which calibration was performed (not the actual one) and the respective calibration values 0% air-sat., temp. at 0% and 100 % air-sat., temp. at 100%. 6. Now, user-defined calibration is complete. Confirm the calibration values by clicking the ‘Finish’ button. When doing so, a message window opens and informs you that you will overwrite the existing calibration values. Click the ‘Continue‘ button to store the new calibration data.