Download MICROX TX3 - Loligo Systems
Transcript
Calibration of Oxygen-Sensitive Microsensors 67 13. Now, calibration with temperature compensation is complete. Confirm the calibration values by clicking the ‘Store’ button. 14. Pull the glass fiber with its sensor tip back into its protective glass housing before removing the microsensor from the calibration vessel. 7.3.5 Manual Calibration A manual calibration should be applied, if you don't want to calibrate your sensor again. However, this is only possible if you already know the calibration values of the special sensor. 1. Connect the Microx TX3 via the RS232 cable to your computer. 2. Switch on the Microx TX3 oxygen meter 3. Start the Microx TX3 software on your computer and click the Calibration menu item. 4. Select the calibration routine ‘calibrate manually’ and click the manual button