Operating Manual SEFRAM 4451 SETUPS Menu F1: RECALL - Recalls a previously stored front-panel setup from the selected buffer. Change the buffer number by using the rotary input knob. Valid storage buffer numbers are from 1 to 99. Buffer 0 is the factory default setup; buffer 99 is the last front panel setup before power-off. F2: STORE - Stores the current front-panel setup to the specified storage buffer. Change the buffer number by using the data keys or the rotary input knob. Valid storage buffer numbers range from I to 98. F4: CLEAR ALL - Clears all data on all memory settings, after a YES or NO selection message. 3.6.6 UTIL Menu Utility Menu F1: GPIB M4451 A00 -Selects the GPIB remote mode of operation. After selection the GPIB can be set to any value from 1 to 31 using the rotary knob. The value is kept in a nonvolatile memory and used at power-on. The factory default address is 10. Setting the address to 31 puts the device in the off-bus state (it will not respond to messages on the GPIB bus). 29