Download DDI Prodigy Pocket Specifications
Transcript
Medtronic Confidential 9879EN-13_9501043EN/CTC.qxd 5/10/99 11:29 Page 13-6 7 x 9 inches (178 mm x 229 mm) 13-6 Understanding Pacemaker Operations Bipolar Sensing Polarity Confirmation Before programming from unipolar to bipolar sensing, the programmer verifies the presence of a functioning bipolar lead by testing impedance for each lead at an Amplitude of 5.0 V and a Pulse Width of 1.0 ms if the permanent settings are at or below this level. If permanent settings are above these values, the measurement will be made at the permanent settings. If lead impedance is between 200 Ω and 3000 Ω, a functional bipolar lead is assumed to be present. If the impedance is outside this range, a functional bipolar lead is not present; the programmer warns that the test failed, and Sensing Polarity remains set to unipolar. This interlock feature may be overridden and lead Sensing Polarity forced to bipolar. WARNING: Do not override for an implanted lead. The override feature is intended only as a means to program the pacemaker for bipolar polarity before lead connection. Determining Sensing Threshold(s) at Implant At implant, a Medtronic Pacing System Analyzer (PSA) may be used to determine sensing threshold values for all Prodigy models except the Prodigy VDD model. Refer to the PSA manual for detailed instructions. (Refer to the CapSure VDD Model 5032 lead technical manual for information on determining sensitivity.) Prior to lead connection, measure the sensing potentials in the unipolar and (when appropriate) the bipolar configurations. When using a bipolar system, an adequate intracardiac signal should be present in both configurations to ensure proper sensing in either. Verifying Sensing Threshold(s) at Follow-up Intracardiac signal amplitudes decrease during the lead maturation process. Medtronic programmers provide an automatic sensitivity test that allows the follow-up clinician to verify a patient’s sensitivity setting(s). The automatic test provides for atrial or ventricular monitoring. The test provides the sensitivity setting just above and below the point at which P or R waves are sensed. For a detailed description of the test procedures, refer to “Determining a Setting for Sensitivity” in Chapter 7. 196407001 Rev A Refer to the “Reference man.” category in doc#163256 for Printing Instructions.