Download Simplicity 1693407 Specifications
Transcript
Model 50 Valve, Continued Step A5E31405710001 Procedure 6. Repeat steps 3 and 4 with the middle plate. 7. Place the top plate on the valve, verifying alignment using the alignment marks. 8. Install the 5 screws and washers finger tight. 9. Tighten the screws down evenly (2.5mm Allen wrench) to 6 to 8 inch-pounds of torque. (It is recommended to use the torque wrench available from Siemens, PN 1631005-001, which is calibrated at 7.2 inch pounds). Remove the assembled valve from the valve fixture. 10. Reinstall the valve in the Analytical Module and connect all tubing. 11. Check for leaks. 12. Reinstall the Analytical Module in the Modular Oven as described in the section of this chapter relating to the Analytical Module. Verify valve operation by running chromatograms. 4-25