Aerosol Management During sample acquisition and sorting, the sort block door and flow cell access door should be kept closed to help contain potential aerosols (Figure 1-14). Cell sorters that use droplet generation methods, like the BD FACSAria II, can produce aerosols around the sample stream. When acquiring biohazardous samples, follow universal precautions at all times. Keep the sort block door closed during sorting. If you need to access the sort block and you are working with highly infectious samples, consider turning off the stream before opening the sort block door. Figure 1-14 Sort block with door closed sort block deflection plates warning light sort collection chamber Additional aerosol removal is provided by the BD™ Aerosol Management Option (AMO). The AMO is an optional device that evacuates the sort collection chamber during sorting. The AMO is equipped with a 0.01-µm pore ultra-low penetrating air (ULPA) filter to trap aerosolized particles. See Appendix C - BD Aerosol Management Option on page 281 for complete information on using this option. Chapter 1: Cytometer Components 31