Figure D-2 Input and output ports on water bath 5 Connect the insulated hoses from the recirculating water bath to the cytometer control panel. NOTE Make sure you match the flow direction from the water bath port to the flow direction on the instrument manifold. The instrument ports are labeled “cooling in” and “cooling out.” Figure D-3 Connections for temperature control option Cooling In 6 Cooling Out Fill the water bath with distilled water containing 0.1 g/L sodium carbonate. Sodium carbonate helps reduce corrosion. Refer to the water bath manufacturer’s documentation for fill levels and other setup information. 7 304 Plug in the water bath power cord. BD FACSAria II User’s Guide