DC Power Supply Operation
Square-Wave Operation
3
Table 3-5 Square-wave amplitude minimum and maximum values
Square wave amplitude value
Range
NOTE
Minimum
Maximum
S1 (30 V/1 A)
0V
30 V
S2 (8 V/3 A)
0V
8V
S1S2 (autoranging)
0V
30 V
• The frequency can be stepped through multiple steps from 0.5 Hz, 2 Hz,
5 Hz, …, 4800 Hz (or 10 Hz to 4800 Hz if range S1S2 (autoranging) is
selected).
• If range S1S2 (autoranging) is selected, the duty cycle is fixed at 50%.
• The duty cycle can be stepped through 256 steps, where each step is
0.390625% more than the previous step. The best resolution the display
can offer is 0.001%.
• The pulse width can be stepped through 256 steps, where each step is
1/(256 × frequency) more than the previous step. The display will
automatically adjust to a 5 digit resolution (ranging from 9.9999 ms to
999.99 ms)