Download Agilent Technologies 1912A Technical data

Transcript
12
SYSTem Subsystem
1xeV-DO
The following table shows the power meter presets when <character_data>
is set to XEVDO. Commands not listed are preset according to their
DEFault values (for further information refer to Table 12- 26).
Table 12-57 1xeV-DO: Power Meter Presets
Command
Setting
Comments
+1900.000 MHz
Frequency setting
NORM
Measurement mode
E9321A/25A: DEF
E9322A/26A: DEF
E9323A/27A: HIGH
N1921/2A: LOW
Sensor video bandwidth
[SENS[1]]|SENS2:SWE[1]|2|3|4
:OFF:TIME
Gate 1: 10 µs
Gates 2 - 4: 0
Delay between trigger point and time
gated period.
[SENS[1]]|SENS2:SWE[1]|2|3|4
:TIME
Gate 1: 810 µs
Gates 2 - 4: 0
Length of time gated period for time
gated measurements.
TRIG[:SEQ[1]|2]:SOUR
INT1
INIT:CONT
ON
Trigger source set up and acquisition
mode continuous triggering
TRIG[:SEQ]:LEV:AUTO
ON
Enable automatic setting of the trigger
level
TRIG[:SEQ]:LEV
AUTO
Power level
TRIG[:SEQ]:SLOP
POS
Trigger event recognized on the rising
edge of a signal
TRIG[:SEQ]:DEL
0s
Delay between recognition of trigger
event and start of a measurement
TRIG[:SEQ]:HOLD
1 ms
Trigger holdoff
OFF
Auto range off
Frequency
[SENS[1]]|SENS2:FREQ[:CW|:FIX]
Sensor measurement mode
[SENS[1]]|SENS2:DET:FUNC
Sensor video bandwidth setup
[SENS[1]]|SENS2:BAND|BWID:VID
Gate Setup
Trigger setup
1
Range
[SENS[1]]|SENS2:POW:AC:RANG:AUTO
568