Calibration Factor Block Layout
E4410 Series
Sensors: Calibration
Factor Block Layout
No.
Bytes
Contents
Data
Format
Data
Range
Units
Notes
A
These table entries are repeated as shown for each frequency point
Frequency (point ‘N’)
4
-
32 bit fixed
0 to Fhpb*
(2^32)
None
Fhbp = Freq Hz per bit
Cal factor (low power)1
2
-
2.14
0.25 to 3
None
Power (in watts) is
divided by this value.
Cal factor (high power)1
2
-
2.14
0.25 to 3
None
Power (in watts) is
divided by this value.
Table Size:
-
See note1
The table size is
dependent on the
number of frequency
points.
Table A-97 Calibration Factor Block Layout: E9300 Series Sensors
E9300 Series
Sensors: Calibration
Factor Block Layout
No.
Bytes
Contents
Data
Format
Number of tables
1
2
Number of frequency
points
2
-
16 bit
integer
Bytes per frequency
point
1
-
-
Frequency LSB weight
2
1000
Data
Range
Units
Notes
None
Number of cal factor
tables. Note that the
power levels and
frequency points are the
same for all tables.
Header:
N1911A/1912A P-Series Power Meters Programming Guide
None
-
None
Number of bytes in cal
factor value at each
frequency and power
level.
-
Hertz
Fhbp (Freq. Hz per bit).
1 KHz per bit for the cal
factor: 1KHz x 2^32 =
4.3E+12 = 4300 GHz
range.