Power Meter Remote Operation
Bit
Number
Meaning
EVENts Causing Bit Changes
11
Lower Limit
Fail
This is a summary bit for the Lower Limit Fail Register.
12
Upper Limit
Fail
•
SET: If a measurement is made and either a channel or window lower limit test
fails.
•
CLEARED: If a measurement is made and the lower limit test is not enabled or
the test is enabled and passes.
1
This is a summary bit for the Upper Limit Fail Register.
•
SET: If a measurement is made and either a channel or window upper limit test
fails.
•
CLEARED: If a measurement is made and the upper limit test is not enabled or
the test is enabled and passes.
Device Status Register
The device status register set contains bits which give device dependent
information.
The following bits in these registers are used by the power meter:
Table 1-11 Bit Definitions - Device Status Register
Bit
Number
Decimal
Weight
Definition
0
-
Not used
1
2
Channel A sensor connected
2
4
Channel B sensor connected*
3
8
Channel A sensor error
4
16
Channel B sensor error*
5
32
Channel A sensor Front/Rear
6
64
Channel B sensor Front/Rear*
14
16384
Front Panel key press
* N1912A only
N1911A/1912A P-Series Power Meters Programming Guide
65