Front Panel Operation
Operation
To Use Couple Mode
1. Press the Couple Mode key to set the couple mode ON. Pressing the key again
sets the mode OFF.
2. Perform the input couple port detection.
a. Press the Menu key. The setup menu will be displayed on the LCD.
b. Move the cursor to SCAN, then press the Enter key. The LCD displays NO.
c. Press the arrow key to display YES, then press the Enter key.
d. Press the Exit key.
3. Change the couple ports (multiple ports can be set).
a. When the couple mode is ON, press the Port Function key. The blink-cursor
appears on the first line of the LCD.
b. Press the right/left arrow key to move the cursor on the input port number
(01 to 14) that will be set to the input couple port.
c. Press the up/down key to display C.
d. Repeat b and c for the all couple ports, then press the Enter key.
To Display Firmware Revision
1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to CONFIG, then press the Enter key.
3. Move the cursor to REVISION, then press the Enter key. The firmware revision
is displayed.
To Display Module Information
1. Press the Menu key. The setup menu will be displayed on the LCD.
2. Move the cursor to CONFIG, then press the Enter key.
3. Move the cursor to UNIT, then press the Enter key. The module information is
displayed.
4. Press the arrow keys to display another module information item.
Agilent B2200 User’s Guide, Edition 4