Download Agilent Technologies B1500A User`s guide
Transcript
Introduction Overview Classic Test You can perform measurements by using the user interface similar to the semiconductor parameter analyzers such as Agilent 4156C. The setup can be made by entering values into the setup tables used for the measurement module control. And it can be saved and recalled as your setup (My Favorite Setup). This measurement environment provides the following functions. • I/V Sweep • Multi Channel I/V Sweep • I/V List Sweep • I/V-t Sampling • C-V Sweep • SPGU Control • Switching Matrix Control • Direct Control The Direct Control test mode is used to control the measurement modules directly by using the B1500A GPIB control commands and perform measurement. For the control commands, see “Programming Guide”. Tracer Test Curve tracer test mode. This test allows you to perform the high speed I-V measurement on one screen. The Tracer Test screen provides GUI for selecting the used channels, setting the sweep output, and displaying the measurement result (tracing the I-V curve). Range of the sweep output and measurement can be changed by the rotary knob on the front panel. Tracer test also provides the oscilloscope view to display the pulse current or voltage output waveform of the MCSMU. You can verify output waveform over the I-V curve. Quick Test You can execute the test setups saved in a preset group (My Favorite Setup) sequentially. Agilent B1500A User’s Guide, Edition 13 2- 7