Calibration 6-3
6-3. Calibration with HOLD active
Press the MODE key.
The legend CALIB
appears, and the YES/NO
key prompt flags flash.
MODE
MEASURE
CAL
DISPLAY
HOLD
YES
NO
MODE
ENT
MODE
HOLD
YES
NO
YES
HOLD
HOLD
YES
NO
HOLD
Put the sensor in standard
solution. Press YES.
ENT
HOLD
After the indication is stable, set the value
ENT
Select the flashing digit with the > key.
Increase its value by pressing the key
When the correct value is displayed,
press ENT to enter the change.
>
>
using the >, , ENT key.
ENT
HOLD
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press YES.
HOLD will be displayed. Press NO to turn off
HOLD and return to the measuring mode.
YES
NO
YES
NO
HOLD
IM 12D08B02-01E