Download Model 42i NOx- User Manual

Transcript
C-Link Protocol Commands
Communications Configuration
Send:
Receive:
Communications
Configuration
set time 14:15
set time 14:15 ok
addr dns
This command reports the TCP/IP address for the domain name server.
Send:
Receive:
addr dns
addr dns 192.168.1.1
set addr dns address
This command sets the domain name server address, where address consists
of four numbers ranging from 0-255 inclusive, separated by “.”.
Send:
Receive:
set addr dns 192.168.1.1
set addr dns 192.168.1.1 ok
addr gw
This command reports the default TCP/IP gateway address.
Send:
Receive:
addr gw
addr gw 192.168.1.1
set addr gw address
This command sets the default gateway address, where address consists of
four numbers ranging from 0-255 inclusive, separated by “.”.
Send:
Receive:
set addr gw 192.168.1.1
set addr gw 192.168.1.1 ok
addr ip
This command reports the IP address of the analyzer.
Send:
Receive:
addr ip
addr ip 192.168.1.15
set addr ip address
This command sets the analyzer’s IP address, where address consists of four
numbers ranging from 0-255 inclusive, separated by “.”.
Send:
Receive:
set addr ip 192.168.1.15
set addr ip 192.168.1.15 ok
addr nm
This command reports the TCP/IP netmask address.
Send:
Receive:
Thermo Fisher Scientific
addr nm
addr nm 255.255.255.0
Model 42i NOx Instruction Manual
B-33