Download Rev AB - AMETEK Programmable Power
Transcript
AC Source SCPI Programming Manual California Instruments trigger the acquisition of new data when they are executed. First, you must initiate the measurement trigger system and generate a measurement trigger as explained in the following section "Triggering Measurements". When the measurement data has been acquired by the voltage and current data buffers for each output phase, use INSTrument:NSELect to select each phase, and FETCh to return the specified measurement data. The following commands return rms voltage: INSTrument:NSELect 1 FETCh:VOLTage:AC?;*WAI INSTrument:NSELect 2 FETCh:VOLTage:AC?;*WAI INSTrument:NSELect 3 FETCh:VOLTage:AC?;*WAI 6.6.6 Returning Voltage and Current Data From the Data Buffer The MEASure and FETCh queries can also return all 4096 data values of the instantaneous voltage and current buffers. These are: MEASure:ARRay:CURRent[:DC]?;*WAI MEASure:ARRay:VOLTage[:DC]?;*WAI 6.6.7 Triggering Measurements You can use the data acquisition trigger system to synchronize the timing of the voltage and current data acquisition with an external trigger source. Then use the FETCh commands to return different calculations from the data acquired by the measurement trigger. The following measurement trigger sources can be selected: IMMediate SYNChronize BUS TTLTrg Aquire the measurements immediate Internal to phase A angle IEEE-488 device, *TRG, or <GET> (Group Execute Trigger) The signal driving the Trigger Out iM Series / i Series / iX Series 149