24 3. Timing 3.2 DiCAM-PRO in Multi Trigger Mode Multi Trigger Mode max. pulse frequency resulting delay between two exposures < 20ns not allowed not allowed ≥ 20ns 2 MHz 500ns Delay Time 0 ... 999ms Exposure Time 20ns ... 999ms) Steps see below! Steps see below! Loops 1 ... 256 The selected delay and exposure time pairs can be repeated up to 256. Each Loop will be triggered separate. (Trigger auto or external) Possible settings for delay and exposure times Following there is an overview of the possible settings for delay and exposure times. Trigger 20ns steps Delay 999ms 0 Exposure 20ns steps 20ns ©PCO 2004 DiCAM-PRO 999ms