1.0
INTRODUCTION
1.1 FOREWORD
Polycold’s Cool Solutions® low temperature gas chillers are used to allow low temperature vacuum
processing of semiconductor device wafers as well as other applications. Vacuum processing systems
use recirculating fluids to keep the wafer at controlled temperature. The wafer sits on a pedestal or table
in a vacuum and is processed, typically in a vacuum with specific gases and metallic vapors to create a
uniform coating on it.
Gas circulating in the secondary loop of the 1XCL closed loop gas chiller is cooled in the gas chiller. The
temperature of the gas leaving the gas chiller can reach temperatures of –100oC. The chilled gas is
recirculated by a secondary gas compressor. The customer-installed lines connecting the gas chiller and
the pedestal are vacuum jacketed to minimize the heat leaks from the ambient to the chilled gas.
The system consists of a refrigeration system, which chills the secondary gas, a secondary gas
recirculating compressor, interconnecting gas lines and the table in the vacuum system. The refrigeration
system uses a propriety and patented blend of refrigerants. The secondary loop compressor uses an inert
gas whose dew point is below –80oC. The refrigeration system as well as the secondary gas loop requires
cooling water. There is no co-mixing of the refrigerant and the secondary gas. The equipment is designed
for continuous operation. Electricity, cooling water, and pressurized dry nitrogen are the only required
utilities.
Pressures are stated as gauge, not absolute. Psig is pounds per square inch gauge and kPa is
Kilopascals gauge.
kPa = 6.895 x Psig
bar = Psig  14.5
This technical manual contains Safety, Service, Introduction, Specifications, Installation, Operation,
Maintenance, and Parts information for all the equipment in the system.
Notes:
REVIEW THE INSTRUCTION MANUAL BEFORE OPERATING THIS EQUIPMENT!
This unit was shipped with a shipping bolt to secure the compressor for transport. It must
be removed to prevent excess vibration. Failure to do so may void your warranty.
Before shipping this unit the shipping bolt must be re-installed to prevent damage during
shipping. Failure to do so may void your warranty. Refer to section 2.3, Item 3.
Upon arrival of the equipment and prior to starting its installation or operation confirm that
the balance pressure (BP) shown on the suction pressure gauge is between 140 – 160 psi.
Refer to section 2.3, Item 2.
825130-00 REV B
Page 4