ExtraLab Plus Cork Extraction Force and Torque Measuring System
3.3.8. Calibration
Like any laboratory equipment, the ExtraLab Plus is subject to calibrations in order to prove and
check the values acquired. If required, EGITRON can supply a Calibration Kit for ExtraLab Plus, as
referred in Annex C.
.
It is not possible to simultaneously perform calibration and tests.
Calibration is an operation with a high level of responsibility and only technical staff with
metrology knowledge should perform this operation. For this reason, the calibration
egitron To increase security this
password should be changed
window,
select
calibration
on the
main menu. Afterwards, it is
necessary
to
type
the
password in order to validate
the user (see note above).
After that,
window will open (Figure 28).
On the upper side of the
window, there is a list for
choosing the load cell, which
will be calibrated. (Extraction
Figure 28 Calibration Window
or Torque)
Below this, there is the load cell value, where is indicated the value obtained thru the Analogue-toDigital Converter (ADC value) and the corresponding value in daN.cm and in Kgf.cm (depending on
the load cell).
At the center there is the Conversion Function stored on the machine memory. This function is a 3rd
degree equation, used to convert the digital value (ADC) acquired from the load cell (converted by
the analogue-to-digital converter) into N (Newton). The coefficients of this function are read from