8. Auto setup
The fifth tab on the Parameters screen
enables the user to set operations to be
performed when the Auto Run button is
selected on the top level screen.
Auto setup enables the user to select
which operations should be added to
the sequence which is executed during
Auto Run. Options are shown in Fig.
20 and include image grab, fringe
detection, editing and various display
and analysis controls.
The use of this feature is described
more fully in Section 22 of this manual.
Fig. 20: Option selection for the Auto Run facility
9. Automatic OPD "Pass / Fail" test
Fringe Master can be used to compare
automatically the P-V and/or rms
measurement of the OPD with user entered
limit values.
Select the "Pass criteria" tab from the
Parameters menu and check PV, rms or
both to enable the check.
Enter the limiting value of the required
value of the PV and/or rms as appropriate.
Quit to the top level screen, and Fringe
Master will display a Pass or Fail marker
each time a new OPD is calculated.
Fig. 21: Selecting PV and rms checks and entering
the limiting values