Lake Shore Model 460 Gaussmeter User’s Manual
Device Specific Commands (Continued)
ALML?
Input:
Alarm Low Point Query
ALML?[term]
Returned: <field value>[term]
Format:
±nnn.nn (Refer to command for description)
Remarks: Use ALMLM? to determine units multiplier.
ALMLM?
Input:
Alarm Low Point Multiplier Query
ALMLM?[term]
Returned: <multiplier>[term]
Format:
a
<multiplier>
u = micro = 10-6, m = milli = 10-3, (blank) = unity, k = kilo = 103
Remarks: Used with ALML? query.
ALMS?
Input:
Alarm Status Query
ALMS?[term]
Returned: <state>[term]
Format:
n
<state>
0 = Off, 1 = On
Remarks: Queries current alarm status. Off means no alarm condition exists, On means an
alarm exists.
ANOD
Default Corrected Analog Out Command
Input:
ANOD <mode>[term]
Format:
n
<mode>
0 = Off (user selected scale)
1 = On (default scale)
2 = Analog output controlled by remote interface (refer to AOCON).
Remarks: Sets default analog output status.
ANOD?
Input:
Default Corrected Analog Out Query
ANOD?[term]
Returned: <mode>[term]
Format:
n
(Refer to command for description)
ANOH
Input:
Analog Out High Setpoint Command
ANOH <field value>[term]
Format:
±nnn.nn
<field value>
Enter sign, 4 or 5 digits, and place decimal point appropriate to range.
Remarks: New value is entered on the same field range as the old value. Setting value to zero first will
change the setting range to present display range.
4-32
Remote Operation