Download Instruction manual
Transcript
Basic Operation Arc Calibration Atmospheric conditions such as temperature, humidity, and pressure are constantly changing, which creates variability in the arc temperature. This splicer is equipped with temperature sensor that is used in a constant feedback monitoring control system to regulate the arc power at a constant level. Changes in arc power due to electrode wear and glass adhesion cannot be corrected automatically. Also, the center position of arc discharge sometimes shifts to the left or right. In this case, the fiber splicing position has to be shifted in relation to the arc discharge center. It is necessary to perform an arc power calibration to eliminate both of these issues. Arc calibration is performed automatically using [AUTO] mode only. So arc calibration does not have to be performed when splicing in this mode. Execute [Arc calibration] before using non-auto mode. When Performing the [Arc Calibration] function change the arc power “factor” value. The factor value is used in the algorithm program for all splicing. The arc power value will not change in the splice modes. Operation procedure 1. Select [Arc Calibration] in [Maintenance Menu 2] to display Arc Calibration screen. 2. Set prepared fibers onto the splicer. Use standard SMF ITU-T G652 fiber for Arc Calibration. Use well prepared fibers for arc calibration. Dust on the fiber surface affects arc calibration. Cleave angle threshold does not link to the parameter "Cleave Limit" in splicing modes. Cleave angle threshold is independently set for arc calibration. See section [Maintenance Settings] to change cleave angle threshold. 3. The ARC Calibration after pressing ENT key. 25 12S_Rev1