Download M-Class Waveform Generator

Transcript
Name
Description
Output ECL Trigger
State Command
Sets or queries the unit VXIbus ECLTn line state. VXI only. The following
considerations apply:
Output ECL Trigger
State Query
•
The ECLTn output source and polarity are selectable.
•
Each ECLTn line state is selected individually.
•
ECLT lines can be sourced and sensed simultaneously.
Command Syntax
OUTPut:ECLTrg<n>[:STATe] <state>
Query Syntax
OUTPut:ECLTrg<n>[:STATe]? Æ<state>
Parameters
Name
Type
Range
<n>
Discrete 0
1
<state>
Discrete ON or 1 Enable
OFF or 0 Disable
ECLT0
ECLT1
Output Event Time
Command
Sets or queries the output event time. This is the length of time that event
driven output pulses will stay high. The time is given in seconds.
Output Event Time
Query
Command Syntax
OUTPut:EVENt[:TIME] <seconds>
Query Syntax
OUTPut:EVENt[:TIME]? Æ <seconds>
Parameters
0004-000074
Name
Type
Range
<seconds>
Float
Time in seconds.
MINimum
50 ns
MAXimum 0.163 s
95