Download Precision LCR Meter
Transcript
PROGRAMMING 0.005~1.275 (5 characters, including a 5mV~1.275 decimal) Example MAIN:VOLT 0.005<^END^M> Query Syntax Returns Auto mode as the current measurement mode. Sets the test signal voltage to 5mV. START MAIN:VOLT?<^END^M>or<^J^M> Starts a measurement in manual mode. Return String <string> Voltage MAIN:VOLT :< variable ><^END> (<variable >= 0.005~1.275) Returns the test voltage. Query Example LCR-800 User Manual MAIN:VOLT?<^END^M> MAIN:VOLT 0.005<^END> Command Syntax MAIN:STAR<^END^M>or<^J^M> Example MAIN:STAR<^END^M> Starts the measurement RANGE HOLD Returns the test voltage (5mV) AUTO/MANU Command/Query Turns range hold on or off or queries the range hold status. Command/Query Syntax MAIN:R.H.:<string><^END^M>or<^J^M> Parameter Sets automatic or manual measurement mode. <String> Range hold Syntax OFF. Off ON.. On MAIN:TRIG:<string><^END^M>or<^J^M> Parameter <String> Test mode AUTO Automatic mode MANU Manual mode Example Query Syntax Example Turn range hold off Query Syntax MAIN:R.H.?<^END^M>or<^J^M> MAIN:TRIG:MANU<^END^M> Return String Sets the measuring mode to manual <string> Range Hold status MAIN:TRIG?<^END^M>or<^J^M> MAIN:R.H.:OFF.<^END> Range hold is off MAIN:R.H.:ON..<^END> Range hold is on Return String <string> Voltage MAIN:TRIG:AUTO<^END> Returns automatic mode MAIN:TRIG:MANU<^END> Returns manual mode Query Example MAIN:R.H.:OFF.<^END^M> Query Example MAIN:R.H.?<^END^M> MAIN:R.H.:ON..<^END> Returns the Range Hold status (On) MAIN:TRIG?<^END^M> MAIN:TRIG:AUTO<^END> 95 96