Download Precision LCR Meter

Transcript
PROGRAMMING
LCR-800 User Manual
Query Example
C.V
Command/Query
Returns the Bias Status (External Bias is on).
Turns Constant Voltage on or off. Queries the constant voltage
status.
Syntax
MAIN:BIAS?<^END^M>
MAIN:EXTB:ON..<^END>
INT.B
MAIN:C.V.:<string> <^END^M>or<^J^M>
Command/Query
<String>
Constant Voltage
Sets and queries the internal bias.
MAIN:INTB:<string><^END^M>or<^J^M>
Syntax
OFF.
Off
Parameter
ON..
On
<String>
Internal Bias
MAIN:C.V.:OFF.<^END^M>
OFF.
Off
Turns Constant Voltage off
ON..
MAIN:C.V.?<^END^M>or<^J^M>
Example
On
MAIN:INTB:OFF.<^END^M>
Parameter
Example
Query Syntax
Turn Internal Bias off.
Return String
<string>
Constant Voltage status
Query Syntax
MAIN:C.V.:OFF.<^END>
MAIN:C.V.:ON..<^END>
MAIN:C.V.?<^END^M>
Query Example
MAIN:C.V.:OFF.<^END>
Constant voltage is off
Constant Voltage is on
Return String
<string>
Query
Returns the Internal Bias status (Off).
Queries the Bias status.
Query Syntax
Internal Bias Status
MAIN:INTB:OFF.<^END>
Off
MAIN:INTB:ON..<^END>
On
MAIN:INTB?<^END^M>
Query Example
MAIN:INTB:OFF.<^END>
Returns the Constant Voltage status (Off)
BIAS
MAIN:INTB?<^END^M>or<^J^M>
EXT.B
MAIN:BIAS?<^END^M>or<^J^M>
Command/Query
Sets and queries the External Bias.
Return String
<string>
Bias Status
Syntax
MAIN:INTB:ON..<^END>
MAIN:INTB:OFF.<^END>
MAIN:EXTB:ON..<^END>
MAIN:EXTB:OFF.<^END>
Internal Bias is on
Parameter
Internal Bias is off
<String>
External Bias
External Bias is on
OFF.
Off
External Bias is off
ON..
On
97
98
MAIN:EXTB:<string><^END^M>or<^J^M>