Download USER MANUAL - Loligo Systems
Transcript
Planar Oxygen Minisensor 16 5.2.4 Oxygen Probe for Inline Measurements in Fermenters (OIM) OIM consists of a fitting made from stainless steel. The oxygen sensor is integrated in the top of the metal fitting (as shown below). The metal fitting is connected to the instrument via a polymer optical fiber. The standard fiber cable length is 2.5 m. OIM is available in different sizes (12 mm, 25 mm) and standard OIM fits to B. Braun Biostat B and B. Braun Biostat C fermenters. Polymer optical fiber Steel housing (diameter: 12 mm) DO Probe immobilized on a glass support This chemo-optical DO probe has outstanding properties: The system can be used after autoclavation without recalibration The system is fully autoclavable up to 100 times In contrast to classical oxygen electrodes, membrane cleaning and frequent replacement is not necessary There are no electrolyte solutions to poison or replenish No time for polarization is needed Long shelf-life measuring range: 0-250 % a.s. limit of detection: 0.15 % a.s. Please note: The OIM is only ready for use together with the Optical Exchange Cap (OEC, see page 17) containing the oxygen sensor membrane. Ordering information OIM D12 / L L Length of the Oxygen Probe for fiber cable Inline Measurement - 2.5: 2.5 m Steel tube - 5: 5 m Diameter [mm] / Length [mm] 12 / 120 12 / 160 12 / 215 12 / 325