12
12.2 Job Menu
This menu contains buttons that define the operation that should be performed. The job is not executed immediately. These buttons only select
which job will be run when the “start button” is pressed. For more detailed
descriptions of the jobs see Chapter 8 and Chapter 9.
I/Phi
Select measurement of flux to current characteristics.
Noise
Select white noise measurement. This gives less information, but is
faster than a measurement of noise spectrum. It also provides the
level of line frequency disturbance and the dc-levels of all the channels.
FFT
Select FFT measurement that gives the spectrum of the signals. This
gives information about the noise as a function of frequency, but it is
slower than a measurement of white noise only.
Signal
Measure signal samples.
Tune
Select automatic tuning of all operating point parameters. This version uses larger steps than the fine tuning (below). Faster than the
fine tuning but not as robust.
Tune (fine)
Select the automatic tuning of all operating point parameters using
small steps.
Tune (gate)
Select automating tuning of gate settings only.
Tune (offs)
Select automatic adjustment of offset values.
Manual tune
Select manual tuning mode.
Meas. derivatives
Select measurement of operating point derivative information.
Meas I-V curve (Level 2 only)
Select measurement of current versus voltage curves.