Download User manual
Transcript
3.6 - Calibration 10. Evaluate the calibration as per section 3.6.6.5 on page 120 4-20mA MODEMPRINTER PWR ON/OFF 100|% F | e | e | d | b|Constants a 50|k0 = -2.18 c|k1 = 14.56 k|k2 = 0.00 | | | 0|_________________ 1.00% 4.00% <----- Span ----> SPC-1000 Communication Terminal EJECT MEMORY CARD Menu 11. Transfer the calibration to the transmitter. Press Correlation r2 = 0.98 %scale span Send data . Check that all data that is displayed is correct. If a code has been entered (see section 3.9 on page 140), key this in correctly and press Yes to transfer all the data to the transmitter. If the message “Transmission fail” appear there is a communication problem. Refer to the Service manual and section 4, item 5. 12. Press Trend in the main menu. Check the stability of the signal over time. If necessary, go back to the main menu and set a suitable time constant. Press Damp 1s (see section 3.6.7 on page 125). Then transfer the new data to the transmitter again. 13. Trim the controller. See section 3.14 on page 150. 14. Document the calibration either manually or by making a printout - refer to section 3.6.6.2 on page 112. Refer also to " Calibration is now complete but can generally be improved by entering additional calibration points, i.e. multi-point calibration. OI218.56/3en MEK-2300 Operating instructions © BTG 2004 107