8101/8102 User Manual Preparation: 1. Configure the Oscilloscope as follows: Input Impedance: 50 ohms Range: 100mV 2. Connect the 8102 Channel 1 output to the Oscilloscope input. Terminate the 8102 output at the Oscilloscope input with the, 50Ω, 20dB Feed through termination 3. Configure the 8102 as follows: Frequency: 100MHz Ch1 Output: On Adjustment: 4. Adjust CAL:SETUP81 to get the signal of 6 divisions on the screen. (Setup 50) Frequency Flatness – Modulation Equipment: Oscilloscope, BNC to BNC cable, 20dB Feedthrough attenuator Preparation: 1. Configure the 8102 as follows: Function: Modulation ON Modulation: Sweep Start Freq: 1MHz Stop Freq: 100MHz Sweep Time: 1ms Marker: 1MHz Amplitude: 6V 2. Connect the 8102 Channel 1 output to the oscilloscope input. Use 20dB Feedthrough attenuator at the oscilloscope input 3. Set oscilloscope input impedance to 50Ω 4. Set oscilloscope vertical sensitivity to 0.1V Adjustment: 5. Adjust C1016 for the best flatness. 7-28