8101/8102
User Manual
Preparation:
1. Configure the Oscilloscope as follows:
Input Impedance: 50 ohms
Range:
100mV
2. Connect the 8102 Channel 1 output to the Oscilloscope input.
Terminate the 8102 output at the Oscilloscope input with the,
50Ω, 20dB Feed through termination
3. Configure the 8102 as follows:
Frequency:
100MHz
Ch1 Output:
On
Adjustment:
4. Adjust CAL:SETUP81 to get the signal of 6 divisions on the
screen.
(Setup 50)
Frequency Flatness – Modulation
Equipment: Oscilloscope, BNC to BNC cable, 20dB Feedthrough
attenuator
Preparation:
1. Configure the 8102 as follows:
Function:
Modulation ON
Modulation:
Sweep
Start Freq:
1MHz
Stop Freq:
100MHz
Sweep Time:
1ms
Marker:
1MHz
Amplitude:
6V
2. Connect the 8102 Channel 1 output to the oscilloscope input.
Use 20dB Feedthrough attenuator at the oscilloscope input
3. Set oscilloscope input impedance to 50Ω
4. Set oscilloscope vertical sensitivity to 0.1V
Adjustment:
5. Adjust C1016 for the best flatness.
7-28