Download - Bio-Rad
Transcript
LCP Chip and Liposome Capturing Kit Regeneration is accomplished by DNA dehybridization using the following conditions. Injection 2 is optional because it is used to remove the remaining lipid assemblies if the regeneration is incomplete with injection 1. For the first injection, use a freshly prepared 8 M solution of urea in deionized water. Table 10. LCP chip regeneration conditions. Injection 1 2 Reagent Orientation Volume, µl Flow Rate, µl/min 8 M urea* 20 mM CHAPS Vertical Vertical 150 150 30 30 * If urea is not available, inject deionized water. GLC Lipid Kit When using the GLC lipid kit, injecting lipid modification solution (20 mM CHAPS) will regenerate the chip surface. Table 11. GLC chip regeneration conditions. Injection 1 Reagent Orientation Volume, µl Flow Rate, µl/min 20 mM CHAPS Vertical 150 30 19