Serial Commands and Queries n = 2: user parameter set #2 n = 3: factory settings #1 n = 4: factory settings #2 Examples: ÆCONFIG:PARSET:BOOT 1<CR> Å<CR><LF> ÆCONFIG:PARSET:BOOT?<CR> Å1<CR><LF> Refer to Table B-6: User Parameters, on page B-32. Status and Error Reporting READ:FAULT? This command reads the error code (refer to Table C-3 on page C-7). 0 means there are no errors. Example: ÆREAD:FAULT?<CR> Å12<CR><LF> User interlock is activated READ:FAULT:LIST? This command reads a string of up to 16 status codes which are currently active (refer to Table C-3 on page C-7). Entries are separated by blanks. 0 means there are no errors. Example: ÆREAD:FAULT:LIST?<CR> Å12 23 24 25<CR><LF> READ:FAULT:HIST? This command reads the history buffer, which consists of up to 16 status codes activated after the most recent startup (refer to Table C-3 on page C-7). Entries are separated by blanks. Example: ÆREAD:FAULT:HIST?<CR> Å0 12 24 13 12 23 24 25 12<CR><LF> *STB? This command reads the system status byte (refer to Table C-1 on page C-1). Example: Æ*STB?<CR> Å1<CR><LF> Laser emission is present B-11