Download pdf manual
Transcript
The calibration procedure As a first step, the calibration is verified with the reference glass plate. If the FiberScan™ measures a test result over ± 5% of the reference value, a re-calibration is required. On the main screen, move the cursor the to the Set menu item and enter the menu by pressing Then select the CAL menu item and press to start the procedure (illustration 42). . Illustration 42 - Calibration start screen When the metal Measure button is pressed, the FiberScanONE™ will measure the reference glass plate. If the test results are within the tolerance range of ±5% of the reference glass plate, the result is accepted (illustration 43) and the calibration menu is automatically exited after approximately 5 seconds. Illustration 43 - Calibration check successful If the test results are outside the tolerance range of ±5% of the reference glass plate (illustration 44), the user will be asked to perform a new calibration by pressing the key. Please leave the glass plate on the sensor! To abort the calibration procedure press . Illustration 44 - Calibration check failed, re-calibration required CAUTION! During the calibration process do not remove the reference glass plate from the sensor and keep distance from it. 37