Download Data Analysis Software InSight & InSight
Transcript
55 System Alarms To program a System Alarm you must check the box that corresponds to the desired response to each alarm condition (see below). You may send the alarm to a pager, via modem, or set one of four relays (depending on which options are installed in the logger). Column M R1 R2 R3 R4 Output Type Modem Relay One Relay Two Relay Three Relay Four Check the box in the ‘M’ column to report via Modem. Check the box in the R1 through R4 column to report via relay 1-4 (if available). A table of each system alarm and its cause is shown below: System Alarm Condition Low Main Battery Memory Battery Low Slate Memory Slate Memory Full Low Bubbler Pressure U-Sonic Echo Loss (A pulse of sound was sent but no echo was received back) Transducer Ringing U-Sonic Failure AS041 Rev. 10-1-99 Cause Batteries require charging or replacement. Internal memory battery voltage is too low. Change memory batteries. Free slate memory is less than 20% Slate memory is full. Download data and restart program. Bubbler system not developing sufficient air pressure. (Inspect air pump, reservoir and associated tubing assemblies for problem) The echo has been temporarily deflected by a change in site conditions such as floating debris or foam in the channel, wind, etc. Transducer is operating within the deadband. Transducer not plugged in. Cable damaged. Transducer thermal sensor damaged. InSight Data Analysis Software