5 Waveform Acquisition
Pulses per Rotation (Pulse/Rotate)
When the time base is an external clock, you can specify how many pulses of the external clock signal (how
many sampled data acquisitions) correspond to one mechanical rotation (or period). For example, if you set
Pulse/Rotate to 100 pulses, when the record length is 1 kpoint, 10 rotations worth of sampled data will be
acquired. When Pulse/Rotate is set to 1 pulse, each point of sampled data corresponds to a single rotation.
The Pulse/Rotate setting only affects the horizontal-cursor measurement values and how the time axis is
displayed on the screen. For example, if you set Pulse/Rotate to 100 pulses, when the record length is 1 kpoint,
1 div will correspond to 1 rotation. With these settings, if you move the cursor by 1 div, the measured horizontal
value will only increase by 1.
Selectable range for pulses: 1 to 24000
Notes about Sampling Using an External Clock Signal
• You cannot acquire waveforms when the acquisition mode is set to Envelope or BoxAverage.
• You cannot display waveforms in roll mode.
• There is no function for dividing the frequency of the clock signal.
• The time axis cannot be changed. To change the time-axis display range, change the record length, or
zoom in on the time axis.
• The measured time values in cursor measurements and automated measurements of waveform
parameters indicate the number of clock signal pulses. For these measurements, units are not displayed.
• Hard disk recording cannot be used.
• The dual capture feature cannot be used.
• The trigger settings listed below are invalid.
Hold-off, trigger delay, period trigger, pulse width trigger
Maximum Sample Rates for Each Module
If you set the sample rate of the DL850/DL850V to a rate that is higher than a module’s maximum sample
rate, because the data is updated at the module’s maximum sample rate, all the data within the module’s data
update interval will be the same. The maximum sample rates for each module are listed below.
Module
701250/701255
701251
701260/701270/701271/701275
(when measuring voltage)
701261/701262
(when measuring temperature)
701280
701265
720210
720220
720230 (LOGIC)
720240 (CAN MONITOR)
Maximum Sample Rate
Internal Clock
10 MHz
1 MHz
100 kHz
100 kHz
500 Hz
25 kHz
500 Hz
100 MHz
200 kHz
10 MHz
100 kHz
External Clock
9.5 MHz
1 MHz
100 kHz
100 kHz
500 Hz
25 kHz
500 Hz
9.5 MHz
200 kHz
9.5 MHz
100 kHz
Notes about Using the 16-CH Voltage Input Module