STANDARD OPERATING PROCEDURE
No: METH 011
Version: 1
No of pages 8
Date: 05-062011
Title: Soil mineralogy analysis using Bruker D2 PHASER
Diffractometer System
I.
SCOPE AND APPLICATION
The determination of the types and relative amounts of the minerals present in soil (soil
mineralogy) is determined routinely because of its strong influence on soil behavior, its use in
soil classification, and its relevance to soil genetic process¬es. ICRAF Soil-Plant Spectral
Diagnostic Laboratory uses X-ray diffraction (XRD) technique for qualitative and quantitative
assessment of minerals in soils. XRD is the most powerful technique used for analysis of
minerals and offers mineral phase’s identification and quantification. It takes two types of
approaches namely (1) Clay fraction approach; clay fractions are physically separated from the
rest of the rock or material and made into an orientated layer of clay supported by a substrate
and (2) Whole rock approach; sample is prepared into a random powder which helps in analysis
of total amounts and identification of non clay minerals present. These analyses provide
information about the clay minerals present in a sample and also the abundance. This SOP
describes the analysis of soil and rock samples for mineral analysis. It can be applied to all types
of soil samples. XRD has a wide range of applications in geology, material science,
environmental science, chemistry, forensic science, and the pharmaceutical industry, among
others. In Agriculture it has application in qualitative and quantitative analysis of actual
minerals in top soils and sub soils, in classification of soils in terms of weatherable and nutrient
rich minerals to measure soil fertility potential and use of mineralogical data in pedo-transfer
functions to directly predict soil functional properties.
II.
RELATED SOPS
a. PRP 003 Soil sample preparation and handling
b.
III.
PRINCIPLE
Powder diffraction is routinely used as a finger-print identification technique of various solid
materials in the laboratory. It is a high-tech, rapid, cheap and non destructive technique for
qualitative and quantitative analysis of crystalline compounds; about 95% of all solid materials
in the soil are crystalline. When X-rays interact with a crystalline substance or powder, a
diffraction pattern called a diffractogram is produced and can be quantified. Information
obtained from this patterns include phase composition of a sample, types and nature of
crystalline phases (minerals) present, crystal structure, amount of amorphous (OM) content,
microstrain, size and orientation of crystallites. XRD has become an indispensable method for
materials investigation, characterization and quality control. The angle and intensity of the
Page 1 of 10
diffracted beam recorded by a detector forms a diffraction pattern, which provides information
about a sample. The diffraction pattern for every phase is as unique as a fingerprint. Phases
with the same chemical composition can have drastically different diffraction patterns. Use of
the position and relative intensity of a series of peaks is used to match experimental data to the
reference patterns in the database such as the Powder Diffraction File (PDF). The PDF contains
over 200,000 diffraction patterns of crystalline phases. Modern computers have automated
search/match programs that determine what phases are present in a sample by quickly
comparing the ds of the unknown to those of known materials in the International Centre for
Diffraction Data (ICDD) Powder Diffraction File (PDF).
IV.
a.
b.
c.
d.
e.
f.
g.
h.
V.
a.
b.
c.
d.
e.
f.
g.
h.
i.
j.
k.
l.
m.
n.
o.
p.
q.
r.
EQUIPMENT
A D2 PHASER Diffractometer with appropriate software
McCrone micronising mill
Weighing balance
Centrifuge
Oven
Ultra-sonic probe
Vacuum pump
Millipore filtration apparatus vacuum tubing
MATERIALS
Corundum and quartz standard reference material
Standard 55mm sample discs made from PMMA
Innovatek cooler fluid
glass microscope slides
Soft brush
Weighing paper
PH-indicator strips (non bleeding)
weighing paper (4"x4")
mortar and pestle
250 µm mesh sieve
spatulas
carbide or diamond glass scriber or marker pen
Squeeze bottle with distilled water, ethanol and hexane
lab tissues
thermometer
Timer
500 ml beakers
Stiff brush or stop cork no. 43
Page 2 of 10
s. Uniform set of sedimentation cylinders (measuring cylinders) with internal depth of 340 ±
20 mm and capacity of 1 L.
t. Millipore HA, 47-mm, 0.45-micron nominal pore opening cellulose filters
u. Desiccators with desiccator shelf
v. Tongs
w. Hexane
x. Ethanol
y. Deionised water
z. Forceps
VI.
PROCEDURES
a. Sample preparation
i. Whole rock approach procedure (randomly oriented powder)
Using a loading device for McCone Mill load the forty-eight cylindrical agate grinding
elements into the sample cup. This helps produce both line contact blows and planar
shearing, in contrast to the random contact blows of a conventional ball mill.
Tare the weight of the sample cup with agate elements and weigh in 3 grams of soil sample.
Using a wash bottle make to 12 grams with ethanol as a grinding fluid.
Cap the cup and mill with a McCrone Mill for 12 minutes. The unique grinding action of the
mill rapidly reduces particles to submicrometer sizes and mixes for homogenization
required for quantitative and qualitative analytical methods (Approx. 10 µm). A longer
milling time lowers the intensity of the peaks. Making the particles too small can cause
damage to crystal structure or alteration of phases making it difficult to identify or quantify
the actual phases present in the sample.
After milling shake the cup just in case milling had stopped a few minutes before and stayed
in the milling position. Wash out all the milled soil into a 50 ml centrifuge tube with ethanol
making sure to collect as much of it as possible.
Turn on the power switch of the centrifuge and press Stop/Open to release the safety latch
and allow the cover of the centrifuge to be opened. Place the tubes in the centrifuge. Be
sure that the centrifuge is balanced by having opposite tubes filled equally with ethanol; 2,
4, 6, or 8 tubes may be used. Close cover. Centrifuge for 10 minutes at 4000 rpm.
6. Remove tubes from the centrifuge, pry off caps, and pour off the supernatant liquid into
the sink. Be careful that the sediment on the bottom of the tubes is not poured off. (NB: we
loose a little of the fine clays when decanting, if this is of interest then it would be better to
evaporate ethanol in an oven instead of centrifuging. )
Add Hexane to the sample in the ratio of 0.5 ml hexane to 1 g of soil sample. Using a vortex
mixture, mix to re-suspend the sample in hexane and then dry at between 80o C to 105o C
in an oven for one hour.
Page 3 of 10
Using a spatula to loosen any sample that has stuck to the centrifuge bottle, brush the dried
material into from the tube to a 250 µm sieve.
Brush the powder sample through the sieve into a weighing paper. A plastic stop cork can
also be used to push the powder through the sieve. The purpose of the brushing is to obtain
a harmonized powder with many crystallites in random orientations.
Pour the sieved powder sample into a glass vial and cap. The sample is ready for loading and
measurement.
ii. Whole rock approach procedure (randomly oriented mounts)
Pour a fraction or portion of the randomly prepared samples into the well of a low
background sample holder.
Tap the holder on a bench or on the side to help fill and properly pack the sample to avoid
sample displacement which causes peak shifts.
Using a sharp razor, tap the sample surface slowly into either direction pushing excess
sample slowly to the end of the well and finally scrapping it off the holder.
Repeat this procedure until the well is completely filled and the sample surface flat with
excess sample on the disk brushed off.
Correct sample height to the lip is critical since error in height will cause peak displacement.
After measurement the disc can be off loaded and washed with tap water and then re-used.
Concern on trace contamination is not necessary
iii. Clay fraction approach
Using a motor and pestle pound the soil sample in to smaller particles. 2mm sieved samples
can also be used.
Scoop approximately one to three spatulas full of the sample into a 500 ml beaker. Top to
the 500 ml mark with deionised water. The amount to be soaked is subjective to the
amount of clay that will make a thick clay mount on the glass slide.
Disperse the sample with the sonic probe for 5 minutes. Rinse off any soil material on the
tip of the probe with deionised water into the beaker and dry between samples.
Transfer the dispersed sample into a 1 L measuring cylinder; fill to the 20 cm mark with
deionised water.
Let it soak overnight to allow material to settle. If the samples are emptied into the cylinder
half an hour or much earlier in the day, it would be necessary to give the cylinder a shake or
stir with a plunger to mix the contents again. According to Stuart’s law spherical particles
will settle faster below the 5 cm mark while the clay particles which are platy and less than
2 µm in size will remain suspended between the 5 and 20 cm mark.
 The following morning, using a siphon tube siphon the suspension from the interval above
the 5 cm mark into a 500 ml bottle for storage until mounting as an oriented aggregate.
Remember to rinse out the siphon tube with deionised water between samples to prevent
Page 4 of 10

cross-sample contamination. The siphon tube is customed to only collect suspension above
the 5 cm mark.
Oriented aggregate mounts are prepared by the filter-peel technique. The oriented
aggregate mounts force the clay mineral particles, usually plate-shaped phyllosilicates, to lie
flat, allowing the operator to direct the incident X-ray beam down the z axis of the minerals
and to record the diagnostic basal diffractions. It is the z axis that shows the extent of dspacing expansion and (or) contraction indicative of certain clay minerals during subsequent
treatments. These treatments include: air drying, glycolation with ethylene glycol, heating
to 400 C, and heating to 550 C. The material from the randomly oriented powder mount
may be used for this analysis.
Select glass slides to fit the sample holder and label them with a scriber. Rinse and dry the
slides to remove any particles.
Assemble the filtration device with the cellulose filter in place. Shake the siphoned clay
suspension and pour approximately 20 ml the into the assembled filtration device. Exact
amount will depend on the density of the suspension, mineralogy, presence of organics, and
grain size of the particles. If the suspension is too thick don’t put so much into the filter.
If the filter clogs, use the pipette or a syringe with a needle attached to remove the excess
clay suspension from the filtration device. The clay layer on the glass slide should be
sufficiently thick (10 mg/cm2) to ensure infinite thickness for X-rays and also production of
independent relative peak intensities.
As soon as the suspension is completely drawn into the filter, remove the glass funnel and
use a single-edge razor or metal spatula or forceps to remove the moist filter from the
filtering device. If the clay film on the filter is too dry, wet with deionised water before
removing from the filtering device.
Place the membrane filter and clay film center-positioned on the glass slide and then
quickly and smoothly lightly rolled across the slide to transfer the clay film to the slide. Any
hesitation or jerky motion during the transfer may cause rippling of the film and disrupt the
preferred orientation.
Place the mounts on a cold plate to dry in a few minutes. The mounts are ready for X-ray
diffraction analysis. The specimens are subsequently exposed to treatments.
Ethylene glycol treatment
o Organic liquids, primarily ethylene glycol and glycerol, are extensively used as an
auxiliary treatment to expand swelling clays. Whether or not a mineral expands and the
amount of expansion can provide essential supplementary information aiding claymineral identification. Swelling clays include smectites (e.g. montmorillonite,
nontronite, and beidellite), some mixed-layer clays, and vermiculite.
o A vapor treatment is used as it causes fewer disturbances of the sample and less
amorphous scattering of X-rays by excess liquid.
Page 5 of 10
o Pour ethylene glycol to about 1 cm depth in base of desiccator.
o Place oriented aggregate mounts on the shelf of desiccator. Additional shelves may be
stacked if necessary.
o Place desiccator in oven at 60 to 70 C for about 4 hours or overnight. Longer times will
not hurt samples. Do not remove mounts until they are to be run on the X-ray
diffractometer.
 Heat treatments
o Heat treatments at various temperatures are commonly used to help identify clay
minerals by revealing changes in crystal structure spacings or loss of the structure.
Depending on the temperature and the mineral species, these treatments can collapse
the structure by dehydration, or in the case of other minerals destroy the crystal
structures. NB: some of the changes caused by the heat treatments may be temporary,
and partial or complete rehydration may occur during cooling.
o Preheat a hot plate to 300oC.
o Place the oriented aggregate mount on the hot plate using tongs. Leave sample on the
hot plate for not less than one hour at 300o C.
o Remove mount by pulling it forward with the wire hook until the edge of the mount can
be grasped with the tongs. Do not remove mounts until they are ready to be run on the
diffractometer.
o X-ray the sample and repeat the above procedure at 550o C.
iv. Preparation of reference mineral mixtures
Relative intensity ratios (RIR) of pure mineral reference standards are obtained by preparing
a 50:50 mixture of a corundum powder and a mineral rock. This results in simulated
patterns for use in fitting patterns for quantitative analysis
Build a library of relatively pure- reference minerals from handpicked rocks, buying from
commercial or private collectors or extract through clay fractionation.
Using a motor and pestle or a rock crusher for hard rock’s crush the rocks to smaller
particles. Separate on paper the different colours of the sample. It is important to know the
colour or the physical characteristic of minerals so that it is easy to identify which part of
the rock to pick as sample. Please refer to the minerals of the world book by P. Lof for this
clarity. Course impurities in the reference clay mineral standards are removed by dispersing
with deionised water. See method for preparing clay oriented mounts.
Obtain finer rock particles by pounding in a motor with a pestle. Avoid grinding by circular
motion since this may destroy the structure of the minerals.
Weigh 1.5 g corundum (Aluminium oxide powder) powder into a weighing boat taking care
not to contaminate the corundum in the bottle. This can be done by first weighing all the
1.5 g corundum we need in the weigh boats.
Page 6 of 10
Add in 1.5 g of the fine rock sample in the weighing boat containing corundum and mill with
a micronizing mill for 12 minutes.
Separately, mill the fine rock sample for 12 minutes.
Proceed with the method for preparing random powder specimens and mounting
Measure the intensities of the two samples for 1 hour as per the correct measurement
procedures. See measurement section.
b. Sample measurement
Samples are placed in the sample holder of the diffractometer. The sample holder is factory
aligned so that the goniometer centre coincides with the surface of the sample. The sample
holder is mounted with four screws which must not be loosened to avoid the sample
holder’s alignment being lost. It also has a spherical handle used to lift the sample holder
into measurement position.
For inserting the sample
o Switch off the high voltage to open the door
o Open the instrument door by sliding it up
o Drop down the spherical handle of the sample holder lift
o Take the sample and insert it into the sample position of the sample holder
o Lift the sample back into the sample measurement position by pulling up the spherical
handle
o Close the instrument door by sliding it down and lock by putting the high voltage on
The sliding front door of the instrument is routinely opened to insert and change samples
Notes:
o Do not forget to lift the sample back to sample measurement position by lifting up the
spherical handle
o The D2 PHASER difractometer is fitted with fans that provide its interior with fresh
cooling air from the environment. When the front door is opened air flows from inside to
outside the housing through the front door; hence only solid compressed powders and
samples should be inserted into the sample holder to avoid sample being blown into the
face or inside the instrument.
The data is collected using a measurement procedure that will ensure high
peak/background ratios, low sensitivity to sample density variations, and a good counting
statistics, while the strongest peaks of 0.334 nm quartz stay within the linearity range of the
counter, and the beam stays within the preparation for the lowest 2theta angle, from which
the analysis will be performed. These conditions will be met by using a Bruker AXS, D2
phaser diffractometer system operating at 30 kV and 10 mA using Ni-filtered Cu-Kα
radiation, at variable rotation of 15o/Min in the angular measurement range of 3 to 75 o
2Theta with an accuracy of 0.02o throughout the measurement range, at 0.5 sec/step to
obtain the X-ray diffractogram.
Page 7 of 10
The intensity of diffraction of a given mineral in the sample under investigation will
continuously be recorded as the sample and detector rotates through their respective
angles as a peak
Recommended settings for Bragg-Brentano systems to guarantee an optimum resolution at
maximum intensity are as follows:
Instrument parameter
Value
Radiation Anode
Cu (Cu
Generator settings
30 kV, 10mA (recommended)
Divergence slit
0.6mm
Axial Soller slit module
Primary 2.5mm, Secondary 2.5mm
Anti scatter screen
1.0mm
Monochromatisation
filter
= 0.154060nm)
kβ 0.5 Ni filter
50 detector opening and 0.5% Ni-filter
Upper and lower discriminator window at 0.240o and 0.160o
Linear LYNXEYE detector
For LynxEye detector all reflections are measured in one uninterrupted range. Other
recommended settings for Cu radiation are as follows:
i) Method name: Quartz reference standard
Measurement parameter
Rotation
Scan type
Two Theta (2θ)
Time per step (s)
Value
15 rpm
Locked coupled (Coupled θ/2θ scan)
1 range start 250, increment 0.0040, stop 270
0.500s
ii) Method name: Soils and clay scans
Measurement parameter
Rotation
Scan type
Two Theta (2θ)
Time per step (s)
Value
15 rpm
Locked coupled (Coupled θ/2θ scan)
1 range start 30, increment 0.0020, stop 750
0.500s
Page 8 of 10
Run time
30 minutes
iii) Method name: Reference samples scans
Measurement parameter
Value
Rotation
15 rpm
Scan type
Locked coupled (Coupled θ/2θ scan)
Two Theta (2θ)
1 range start 30, increment 0.0020, stop 750
Time per step (s)
3s
Run time
3 hours
Close door and switch on HV to set generator to 30kV and 10mA.
Initialize 2θ and Phi motor using the initialize button on the commander screen.
Entervalue for sample rotation in rpm as per methods above and press ‘set’ button to start
rotation.
Select scan type, scan mode, increment and time as per method being run. Start
measurement by pressing the start button.
After the measurement is complete save the raw scan file in the appropriate folders of the
D2 PHASER computer.
c. Shutting Down The Instrument
VII.
CALCULATIONS
The software application used to run the analyzer (?????) automatically stores data acquired
during an analysis in comma-delimited ASCII text format for later computation, display, and
printing.
VIII. QUALITY ASSURANCE/QUALITYCONTROL
Reference material:
Blanks:
Calibration:
Duplicates:
IX.
DATA VALIDATION
a. Data Review
Check the sample data and QC data to verify that the data to be reported is based on
acceptable analyses and meet acceptable limits.
Verify that all standard, reagents, and quality control samples have been entered into the
appropriate logs.
Verify that the results have been entered into the appropriate spread sheet and proof read.
Place data in the laboratory database.
b. Supervisor Review
Page 9 of 10
Check the sample data and QC data to verify that the data to be reported is based on
acceptable analyses and meet acceptable limits
Verify that the results have been entered into the appropriate spreadsheets correctly.
If any errors are found, return data to analytical chemist for corrections.
Ensure documentation on printouts (i.e., initials, date, label) is complete.
Authorize reporting.
X.
HEALTH AND SAFETY
Personal protection: Safety glasses and protective gloves are recommended whenever
reagents or samples are handled. For other precautions and safety procedures, consult the
Material Safety Data Sheets (MSDS) for each chemical used.
Electrical hazards: Electrical systems must conform to the ICRAF standards. Shock hazards
exist inside the instruments. Only an authorized service representative or an individual with
training in electronic repair should remove panels or circuit boards where voltages are
greater than 20 V. The instruments require a third-wire protective grounding conductor.
Three-to-two wire adapters are unsafe for these instruments.
Radiation protection: To ensure maximum radiation protection, the S2 Picofox must be
operated with the control software ‘SPECTRA’ included in delivery.Only this software
guarantees the compliance of the X ray source operation parameters as certified according
to anti-radiation precautions. Further ,safety rules and technical installation and
information provided in the user manual should adhered to explicitly.