Download MTX 1052/1054 - Chauvin Arnoux Group
Transcript
Applications Applications (contd.) 8. Measurement in "Analyser" mode Reminder Initially, a frequency signal between 40 Hz and 1 kHz should be injected on channels CH1, CH2, CH3 or CH4. - Only CHx channel signals (and not the Mathx functions) can be the subject of harmonic analysis. - In Analyser mode, the time base is not adjustable. • Set the amplitude of the channels in Oscilloscope mode correctly signals displayed should not be saturated). • On the Instrument menu, select Analyser or click on the (the icon on the toolbar. Reminder The harmonic content of the signal for channels CH1, CH2, CH3, CH4 is represented by "full" bars in the colour of the channel (red for CH1, green for CH2, blue for CH3 and pink for CH4). • The SIGNAL box under the breakdown indicates: - the active channel(s) - the RMS voltage of the signal in Volts - harmonic distortion rate (in %) of the signal • The Reference box enables the reference harmonic to be selected for the measurements. VII - 132 Virtual digital oscilloscopes