MC151 Current calibrator
Powertek
If calibrator is out of limits is in some points of this test, appropriate function and range should be
recalibrated. It is not necessary to recalibrate all functions, but only this one, which does not meet
specification. See chapter Calibration mode, where recalibration procedure is described.
Tables of limits
1 A DC Basic range with linearity test
Function
Range
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
Table I
1
1
1
1
1
1
1
1
1
1
1
1
1
1
A
A
A
A
A
A
A
A
A
A
A
A
A
A
Value (A)
Frequency (Hz)
Deviation
allowed (µA)
200
225
250
275
300
325
350
200
225
250
275
300
325
350
Frequency (Hz)
Deviation
allowed (µA)
105 µ
105 µ
700 µ
700 µ
1.75 m
1.75 m
4.5 m
4.5 m
15.0 m
15.0 m
30.0 m
30.0 m
60.0 m
60.0 m
Frequency (Hz)
Deviation
allowed (µA)
105 µ
350 µ
500 µ
800 µ
1.75 m
4.5 m
15.0 m
30.0 m
60.0 m
0.4
0.5
0.6
0.7
0.8
0.9
1.0
-0.4
-0.5
-0.6
-0.7
-0.8
-0.9
-1.0
DC current test
Function
Range
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
A-DC
Table II
300.0 mA
300.0 mA
2.0 A
2.0 A
5.0 A
5.0 A
10.0 A
10.0 A
30.0 A
30.0 A
60.0 A
60.0 A
120.0 A
120.0 A
Value (A)
0.3
-0.3
2.0
-2.0
5.0
-5.0
10.0
-10.0
30.0
-30.0
60.0
-60.0
120.0
-120.0
AC current test
Function
Range
A-AC
A-AC
A-AC
A-AC
A-AC
A-AC
A-AC
A-AC
A-AC
Table III
300.0 mA
1.0 A
1.0 A
2.0 A
5.0 A
10.0 A
30.0 A
60.0 A
120.0 A
35
Value (A)
0.3
1.0
1.0
2.0
5.0
10.0
30.0
60.0
120.0
55
55
800
55
55
55
55
55
55
Operation manual