Download Report 2 - CSA Group
Transcript
S ECTION 4 SCREENS AND SETTINGS • Filter loading: This field contains the filter loading percentage (Section 5). It indicates the portion of the TEOM filter’s total capacity that has been used. The instrument will always show a nonzero value even if no filter is mounted in the mass transducer. New filters generally exhibit filter loading percentages of 15% to 30% at a flow rate of 3 l/min, and less at lower flow rates. TEOM filter cartridges must be exchanged before this value reaches 100% to ensure the validity of the data generated by the instrument. At some point above 100%, the main flow drops below its set point. If the filter loading percentage is higher than 30% when a new TEOM filter is placed on the mass transducer, or if the lifetime of consecutive TEOM filter cartridges becomes noticeably shorter, inspect the in-line filters for the PM-2.5 and coarse flow lines (Section 5) and replace them, if necessary. In addition to the information indicated above, the instrument may be configured to calculate equivalent MC values for the U.S. EPA or other agencies. For U.S. EPA PM-2.5 reporting, this configuration must be selected and data recorded. Refer to Section 4 for instructions to configure this setting. If this configuration is not selected, this column will not appear on the display. 4-4 OPERATING GUIDE, TEOM 1405-DF THERMO FISHER SCIENTIFIC