Table 7-3 Run Window Settings for the ABI 373 with XL Upgrade
Data Field
Setting
Plate Check Module
Plate Check
PreRun Module
PreRun
Run Module
GS Run
Collect Time
4 hours for a 12-cm gel, 8 hours for a 24-cm gel, or
12 hours for a 34-cm gel
Note
The default collect time is 12 hours.
Sample Sheet
The name of the sample sheet that you created for
this run
Well-to-Read distance
12, 24, or 34 cm
Filter Set
A
Lanes
As appropriate for the comb used
Run Mode (XL only)
Full scan: 24, 34, or 36 lanes
XL scan: 50 or 66 lanes
Operator
Your name
Gel’s Matrix File
The name of the matrix file that you created for this
set of run conditions or None if you have not yet
created a matrix
AutoAnalyze
On or off as desired
Note
You must create a matrix file before
analyzing the data.
continued on next page
SSCP Analysis 7-15