Download xs series manual
Transcript
XS series operating manual 3 Megohmmeter function (insulation resistance measurement) 500V megohmmeter function Measurement voltage Voltage adjustable from 10V to 500VDC by 1 volt step. Accuracy : ±(1% + 1V). Mise en forme : Puces et numéros Positive pole of the HV generator is grounded ( earth) Dynamic stability for a mains ∆V = ±10% : >1%. The maximum current in the measurement circuit is : 2 mA ±20%. The discharge current is limited by a 2.2 kΩ resistor. Mise en forme : Puces et numéros Measurement range Display : 2000 digits The measurement range is defined by the following formula : (Utest / Umax generator ) x 200 GΩ * * or 2TΩ according to the option This gives the following board for the 50,100, 250, 500VDC standard voltages 200GΩ basic version Voltage Measurement range 50V 50kΩ to 20GΩ 100V 100kΩ to 40GΩ 250V 250kΩ to 100GΩ 500V 500kΩ to 200GΩ 2TΩ option version Voltage Measurement range 50V 50kΩ to 200GΩ 100V 100kΩ to 400GΩ 250V 250kΩ to 1TΩ 500V 500kΩ to 2TΩ Measurement accuracy 2000 digits numeric display with unity indication (KΩ, MΩ, GΩ,TΩ). Accuracy (% of the reading, 1U = 1 digit ) : 200 GΩ basic version 2 TΩ option and U<=100VDC 2 TΩ option and U > 100VDC - 17 - DXS and SXS models ±(1.5% + 1U) ±(2% + 1U) ±(1% x Utest / 100 + 1U) Mise en forme : Puces et numéros