Download ROTORS AND TUBES
Transcript
Using Vertical Tube and Rack-Type Rotors REMOVAL AND SAMPLE RECOVERY ! CAUTION If disassembly reveals evidence of leakage, you should assume that some fluid escaped the container or rotor. Apply appropriate decontamination procedures to the centrifuge, rotor, and accessories. 1. Turn the tie-down knob to the left (counterclockwise) to release the rotor from the drive spindle. 2. Lift the rotor straight up and off the drive spindle. If the rotor sticks to the drive spindle, a rotor removal tool may be used. ➠ NOTE Lubrication of the centrifuge drive spindle hub with Spinkote should prevent the rotor from sticking. 3. To remove a tray from a carrier, lift the end of the tray just enough to clear the carrier. Slide the tray up so that it passes under the hinge pins. 6-10 J Series Rotors and Tubes