DSPEC Pro® Digital Gamma-Ray Spectrometer User’s Manual 794380D / 0914 CLEAR_ALL This command is equivalent to the combination of CLEAR_COUNTER, CLEAR_DATA, CLEAR_PRESETS, and CLEAR_ROI commands. CLEAR_COUNTER The live time and true time counters are set to zero. CLEAR_DATA The channels of spectral data in the window of interest (see SET_WINDOW command) are set to zero. The ROI flags are not changed, nor are the presets changed. CLEAR_HV_HEALTH Clears (resets) the state-of-health bits. This command clears the SOH bits in the DIM and does not affect the stored SOH (which is reported when the instrument is not acquiring data). CLEAR_LIST Clears data and resets acquisition in List Mode. The response is the same as the CLEAR command. CLEAR_MDA_PRESET Clears all MCA preset values. This command is used to quickly clear the list of MDA preset values. CLEAR_PRESETS The live time, true time, ROI integral, ROI peak, and overflow presets are all set to zero (disabled). CLEAR_ROI The region-of-interest flags for the channels in the window of interest (see SET_WINDOW command) in the currently selected device (see SET_DEVICE command) are cleared. CLOSE_FILE_FLASH crc No function in the DSPEC Pro; included for backward compatibility. COMP No function in the DSPEC Pro; included for backward compatibility. DISABLE_BLRE_AUTOMATIC Disables the automatic determination of baseline restorer time constant. See also ENABLE_ BLRE_AUTOMATIC, SET_BLRE, and SHOW_BLRE. 56