794380D / 0914 4. FIRMWARE COMMANDS AND RESPONSES INITIALIZE_ZERO_STABILIZATION Resets the zero peak stabilization adjustment to unity (no adjustment). This value is reported as 2048 by the SHOW_ZERO_ADJUSTMENT command. See also SET_ZERO_ ADJUSTMENT, ENABLE_ZERO_STABILIZATION, and DISABLE_ZERO_ STABILIZATION. LIST_BLRE Lists the valid settings of the baseline restore function. See also SET_BLRE. Response: BLRE 10.0 100.0 10 100 Legal settings are 10.0 to 100.0 in steps of 1. LIST_CORR_FLAT Lists the valid settings of the flattop width. See also SET_CORR_FLAT. Response: CORR_FLAT !1.0 .992 0 255 Legal settings are from !1 to 0.992 in (255 !0 + 1 = 256) steps. LIST_DISPLAY Not used by the DSPEC Pro; included for backward compatibility. LIST_GAIN_COARSE Lists the valid coarse-gain settings. Response: GAIN_COAR 1 2 4 8 16 32 LIST_GAIN_CONV Lists legal conversion gain settings. Response: GAIN_CONV 512 1024 2048 4096 8192 16384 LIST_GAIN_FINE Lists the valid fine-gain settings. Response: GAIN_FINE 0.45 1.0 88437 196607 0.45 to 1 in 108171 steps. LIST_HV Lists the valid high-voltage settings. Response: HV 0 5000 0 5000 0 to 5000 in 5000 steps. 61