Download User`s Manual
Transcript
Micro Digital X-ray Processor - Technical Reference Manual mdo-microDXP-MAN-1.1.2 σt = sqrt( σf2 + (1+1/128)σe2) Equation 4-6 This results in less than 0.5 eV degradation in resolution, even for very long peaking times, when resolutions of order 130 eV are obtained. 4.4.2 Applying a Baseline Cut can improve performance when the Baseline Histogram is nonGaussian. Outlying data points are‘cut’ from the running Baseline Average (though still included in the histogram) Raw Baseline Measurement The output of the baseline filter (either the energy filter itself, or a derivative of it) is sampled periodically in the explicit absence of an x-ray step, defined by a baseline threshold. In practice, the DXP initially makes a series of NB baseline measurements to compute a starting baseline mean. It then makes additional baseline measurements at quasi-periodic intervals to keep the estimate up to date. These values are stored internally and can be read out to construct a spectrum of baseline noise, referred to as the Baseline Histogram. This is recommended because of its excellent diagnostic properties. When all components in the spectrometer system are working properly, the baseline spectrum should be Gaussian in shape with a standard deviation reflecting σn. Deviations from this shape indicate various pathological conditions which may also cause the x-ray spectrum to be distorted and therefore have to be fixed. The situation is remedied by removing (“cutting”) outlying samples from the baseline average described below. If the maximum in the baseline distribution lies at E0, then captured baseline values that deviate from E0 by more than ∆E+ and ∆E-, respectively, are not included in the running baseline average. Note that all captured baseline values are included in the Baseline Histogram, however, so that it is always a valid representation of the system’s behavior. 4.4.3 Baseline Averaging in the DXP A running average of baseline measurements is computed, which is then subtracted from sampled peak values to compute the energy of corresponding incident x-rays. The number of baseline samples averaged is set in microManager as “Number of Baselines to Average”. In the DSP this is converted into the parameter BLFILTER according to the equation: # baseline samples averaged = 32768 / BLFILTER Decimation # Baseline Samples to Average BLFILTER (DSP Parameter) 0 2 4 64 128 256 512 256 128 Table 4.1: Typical values used for baseline averaging. The best value for each decimation should be determined empirically, though the general trend illustrated in the table, i.e. larger number to average for higher decimations, should be followed. Physically, the baseline is a measure of the instantaneous slope (volts/sec) for a pulsed-reset detector, and a measure of the DC offset for an RCfeedback preamplifier. For a perfect detector and preamplifier the baseline value is independent of time. In fact, the variation in leakage current of the 8/6/2009 47