Download UTAS user manual - LKC Technologies, Inc.

Transcript
UTAS Software
Note: All flicker stimuli should have a pre-stimulus baseline of 0 in the timeframe
tab
Once all the tabs for the first step of your protocol are checked, click continue, and the choice to
add another step or finish the protocol (which would create a one-step protocol) will come up. If
add another step is selected, the software will automatically use the values selected in the
previous as default.
10.8.2
Editing user-defined protocols
UDPs can edit any user defined protocols by going
to Design Protocol -> Design selecting the
protocol to be edited and clicking on Edit.
The choices given are to add, edit, insert or delete
steps.
10.8.3
Using user-defined protocols
To use the UDP that has been created, go to Perform Test -> User Defined Protocols and select
that protocol from the list.
10.8.4
Changing Standard Protocols
EMWIN comes with a preset number of ISCEV standard protocols such as On/Off, standard,
double flash, etc. It is possible to change those protocols to fit the specific clinic’s needs using
the methods described above.
Version 2.4
71