Download PICOTEST

Transcript
And
[:SENS]:FUNC[:ON] is not Totalize or Voltage Peaks.
See :TRIG:COUN:AUTO for specifics.
Notes
z When :TRIG:COUN:AUTO and :CALC3:AVER[:STAT] are both ON, the
counter clears the statistics results and the current count when it
receives :INIT[:IMM].
z When
:CALC2:LIM:CLE:AUTO
is
ON
and
whenever
the
counter
receives :INIT:CONT ON command, it :
o
Invalidates the limit data,
o
Clears the pass and fail counts (:CLAC2:LIM:FCO and :CALC2:LIM:PCO)
in the limit testing.
o
Sets the Limit-Detect output to the in-limit voltage level.
z When [:SENSe]:EVEN:LEV[:ABS]:AUTO is ON and :INIT[:IMM] command is
sent, the counter issues an auto-trigger on the source.
z If the counter is in the process of making a measurement or
if :iNItiate:CONTinuous is set to ON, an :IMMediate command has no effect,
and error -231 (Init ignored) is generated.
z Beginning a measurement or block of measurements with an :INIT[:IMM]
sets the pending Operation Flag to true. Completing the measurement or
block of measurements (normally or by aborting) sets Pending Operation
Flag to false.
z For an automatically armed Totalize measurement, use :ABORt to terminate
the measurement initiated with :INIT[:IMM].
:INPut[1|2]:ATTenuation 1|10
Sets or queries the input attenuation.
Return
z Numeric data transferred as ASCII bytes in <NR1> format.
Note
z The input attenuation is set to X1 after *RST command.
:INPut[1|2]:COUPling AC|DC
Sets or queries the input coupling.
123