PeakNet-PA Software
• The Alarm Type column lists the types of alarms that can
occur on a channel.
• The CC80 Alarm column assigns alarms to the four alarm
LEDs on the CC80 front panel. With the assignments given
in the following example, if a fatal hardware alarm occurs,
the first CC80 alarm LED lights; if a non-fatal hardware
alarm occurs, the second LED lights, and if a component
alarm occurs, the third LED lights.
• The Relay column lets you open the specified relay (either 1
or 2) when the alarm occurs.
• The Standby and Shutdown columns let you specify to run
the Standby method or shutdown the channel when the
alarm occurs.
• The Bypass column lets you skip injections from a selected
valve when the alarm occurs. For example, if a liquid sensor
on a sample pipe is connected to TTL1 and the pipe is dry,
the sensor sends the signal to trigger the TTL1 alarm.
Injections from that sample are then bypassed.
14. Click on OK to close the Analyzer 1 dialog box, and then close
the Analyzer Configuration dialog box.
2-24
Doc. 031322-02 2/02