Chapter 1 – Performance Specification
Overview of sterilization programs
Type tests
Program type as per
→DIN EN 13060
UniversalProgram
QuickProgram B
QuickProgram S
PrionProgram
Type B
Type B
Type S
Type B
Type B
→Dynamic pressure test of the
sterilization chamber
X
X
X
X
X
→Air leak
X
X
X
X
X
→Empty chamber test
X
X
X
X
X
→Solid load
X
X
X
X
X
→Porous partial load
X
X
X
→Porous full load
X
X
X
→Hollow body B
X
→Hollow body A
X
X
X
X
→Simple wrapping
X
X
X
X
→Multiple wrapping
X
X
X
Drying →solid load
X
X
X
Drying →porous load
X
X
X
X
X
Sterilization temperature
134°C
134°C
134°C
121°C
134°C
Sterilization pressure
2 bar
2 bar
2 bar
1 bar
2 bar
5.5 min.
3.5 min.
3.5 min
20.5 min.
20.5 min.
Sterilization time
X = Conformity with all applicable sections of the standard →DIN EN 13060